• No products in the list
Quotation

The Dermatophagoides pteronyssinus protein Der p 1 is one of the major house dust mite allergens. Citeq Biologics produced a Enzyme-Linked Immunosorbent Assay (ELISA) kit, which can be used to determine the concentration of this allergen in solution in the nanogram range (3,1 – 100 ng/ml). We have developed a kit for five 96-well plates. It provides sufficient antibodies and Der p 1 standard to perform ELISA experiments on five 96-well plates, for a total of 200 samples (performed in duplicates). All components of the kit are supplied in freeze dried form.
■ 1 vial of freeze dried anti-Der p 1 capture mouse monoclonal antibody
■ 1 vial of freeze dried anti-Der p 1 rabbit polyclonal detection antibody
■ 5 vials of freeze dried Der p 1 standard
■ Certificate of Analysis
■ User manual with plate layout

If you have any questions about this kit, please contact us.  

Summary ELISA der p 1 kit

The Dermatophagoides pteronyssinus protein Der p 1 is one of the major house dust mite allergens. Using this Enzyme-Linked Immunosorbent Assay (ELISA) kit the concentration of this allergen in solution can be determined in the nanogram range (3,1 – 100 ng/ml). This kit provides sufficient antibodies and Der p 1 standard to perform ELISA experiments on five 96-well plates, for a total of 200 samples (performed in duplicates). All components of the kit are supplied in freeze dried form.

Principle

Monoclonal antibodies against Der p 1 are coated in the wells of a microtiter plate. This coating captures Der p 1 present in samples and in the standard. After adding the samples, a polyclonal antibody is used to detect the Der p1 captured, afterwards a secondary antibody against the polyclonal is used conjugated with an enzyme.  Conversion of a chromogenic substrate by the peroxidase is quantified after stopping the reaction by measuring the optical density at 450 nm.

Technical data

Specificity: Specific for Der p 1 allergen. Less than 0,5% cross-reactivity with Der f 1 allergen.

Working Range: 3,1 – 30 ng/ml.

Precision: Repeatability (intra-assay precision): 5,1 – 17,5% CV. Intermediate precision (inter-assay precision): 5,8 – 17,5% CV.

Precautions

  • This kit is intended for research purposes only.
  • Upon arrival the kit should be stored at ≤-20°C in a freezer.
  • Protect your plates against contamination by covering them with a lid or parafilm
  • The kit should not be used beyond its expiry date.
  • Wear disposable (latex) gloves when handling specimens and reagents.
  • Use disposable pipette tips throughout the procedure to avoid contamination of reagents.

Contents of the ELISA der p 1 kit

  • Freeze dried anti-Der p 1 capture antibody (1 vial)
  • anti-Der p 1 detection antibody (1 vial)
  • Freeze dried Der p 1 standard (5 vials)
  • Certificate of Analysis
  • User manual with plate layout

ELISA Kit HDM

Additional Materials, Equipment, and buffers

The following materials and equipment are required but are not provided with the kit.

  • Precision pipettors with disposable tips
  • Vortex mixer
  • Tubes for sample and standard dilution
  • 96-well plates suitable for ELISA experiments (for example Nunc MaxiSorp plates)
  • Microtiter plate reader capable of measuring at a wavelength of 450 nm
  • Orbital plate shaker
  • Microtiter plate washer (not necessary, but recommended)
  • Analysis software capable of performing four-parameter logistic regression (not necessary, but recommended)

The following buffers and solutions are required and not provided:

  • Deionized water
  • Phosphate buffered saline, pH 7,4 (PBS)
  • ELISA coating buffer: 0,05 M Carbonate-Bicarbonate, pH 9,6
  • Wash buffer: PBS + 0,05% Tween 20 (PBS-T)
  • Dilution and blocking buffer: PBST-T + 1% BSA
  • TMB substrate solution
  • Stop solution: 0,33 M Sulfuric acid
  • A Goat anti-rabbit antibody conjugated with HRP

Test procedure

Preparation

During step 3 of the assay procedure, reconstitute one vial of Der p 1 standard with 450 µl of dilution buffer. Gently vortex and leave to stand for 10 minutes at room temperature. To make standard 1, gently vortex the vial again, take the volume indicated on the vial (volume x), and add dilution buffer up to a total volume of 500 µl.  We recommend a working range from 30 ng/mL to 0,5 ng/mL. From mastermix, make 2-fold serial dilutions in tubes as indicated in the table below. The Der p 1 standard is intended for single use and should not be stored.

Standard ng/ml Der p 1 Source Added dilution buffer
mastermix 100 x µl of reconstituted standard 500 – x µl
1 30 150 µl from mastermix 350 µl
2 15 250 µl from standard 1 250 µl
3 7,5 250 µl from standard 2 250 µl
4 3,75 250 µl from standard 3 250 µl
5 1,875 250 µl from standard 4 250 µl
6 0,938 250 µl from standard 5 250 µl
7 0,469 250 µl from standard 6 250 µl

Assay procedure Der p 1 ELISA kit

  • Reconstitute the capture antibody in 500 µl PBS. Gently vortex and leave to stand for 10 minutes at room temperature. Gently vortex again just before use. Dilute the capture antibody 100x by adding 100 µl antibody in 10 ml ELISA coating buffer. Immediately add 100 µl to each well of a 96-well microtiter plate to be used in the assay. Incubate for one hour at room temperature on an orbital plate shaker. Alternatively, incubate overnight at 2-8°C.

Optional: For ease of use all of the capture antibody can be coated on five plates in one go and treated with our Coating Stabilizer (article number 100.05.01) after step 4 of the procedure. Treated plates can be stored for up to eight months at 2-8°C.

  • Wash the plate three times with 300 µl PBS-T per well.
  • Add 200 µl of blocking buffer to each well. Incubate for one hour at room temperature on an orbital plate shaker.
  • Wash the plate three times with 300 µl PBS-T per well.
  • Add 100 µl of standards, blank, and samples in duplicate to appropriate wells. A suggested plate layout can be found in the additional information section. Use only dilution buffer for the blank wells. Dilute samples so that at least one measurement falls within the working range of the standard curve. For unknown samples we suggest using 5-fold serial dilutions with eight or more dilutions. Incubate for 30 minutes at room temperature on an orbital plate shaker.
  • Wash the plate three times with 300 µl PBS-T per well.
  • Dilute the detection antibody 300x by adding 33 µl antibody in 10 ml dilution buffer. Add 100 µl to each well. Incubate for 30 minutes at room temperature on an orbital plate shaker.
  • Wash the plate three times with 300 µl PBS-T per well.
  • Add the antibody goat anti-rabbit (not delivered) with dilution buffer in the recommended diluting amount as provided by the manufacturer. Incubate for 30 minutes at room temperature on an orbital plate shaker.
  • Wash the plate three times with 300 µl PBS-T per well.
  • Add 100 µl TMB substrate to each well. Incubate in the dark at room temperature for 5 minutes.
  • Stop the reaction by adding 100 µl stop solution to each well. Gently shake the plate to mix.
  • Measure the optical density with a plate reader at 450 nm within 30 minutes of stopping the reaction.

Calculations

  • Correct the measured OD450 values of the standard and samples by subtracting the mean blank value.
  • Generate a four-parameter logistic curve from the corrected standard values. Use weighting factor 1/Y2.
  • Calculate the sample concentrations by interpolation of the corrected sample values.
  • Correct the concentrations for dilution for samples that fall within the working range of the assay (0,469 – 30 ng/ml) to yield the final sample concentrations.

Additional information ELISA der p 1 kit

Example standard curve

  • Below a typical standard curve that was generated using the BioTek Gen5 software.
  • Do not use this curve in your calculations. Every time an assay is performed a standard curve has to be included.
Curve Formula A B C D R2
Y = (A-D)/(1+(X/C)^B) + D -0,0221 1,24 42,4 2,04 1

 

Suggested plate layout ELISA der p 1 kit

12 Spl 4 Spl 4 Spl 16 Spl 16 Spl 28 Spl 28 Spl 40 Spl 40
11 Spl 3 Spl 3 Spl 15 Spl 15 Spl 27 Spl 27 Spl 39 Spl 39
10 Spl 2 Spl 2 Spl 14 Spl 14 Spl 26 Spl 26 Spl 38 Spl 38
9 Spl 1 Spl 1 Spl 13 Spl 13 Spl 25 Spl 25 Spl 37 Spl 37
8 Blank Blank Spl 12 Spl 12 Spl 24 Spl 24 Spl 36 Spl 36
7 Std 7

0,469 ng/ml

Std 7

0,469 ng/ml

Spl 11 Spl 11 Spl 23 Spl 23 Spl 35 Spl 35
6 Std 6

0,938 ng/ml

Std 6

0,938 ng/ml

Spl 10 Spl 10 Spl 22 Spl 22 Spl 34 Spl 34
5 Std 5

1,875ng/ml

Std 5

1,875ng/ml

Spl 9 Spl 9 Spl 21 Spl 21 Spl 33 Spl 33
4 Std 4

3,75 ng/ml

Std 4

3,75 ng/ml

Spl 8 Spl 8 Spl 20 Spl 20 Spl 32 Spl 32
3 Std 3

7,5 ng/ml

Std 3

7,5 ng/ml

Spl 7 Spl 7 Spl 19 Spl 19 Spl 31 Spl 31
2 Std 2

15 ng/ml

Std 2

15 ng/ml

Spl 6 Spl 6 Spl 18 Spl 18 Spl 30 Spl 30
1 Std 1

30ng/ml

Std 1

30ng/ml

Spl 5 Spl 5 Spl 17 Spl 17 Spl 29 Spl 29
A B C D E F G H

 

If you want to place a direct order of a small vial, please visit our sister company Alfateq.